Annual Calibration Procedures for Masterflex Pumps

7 min read
Updated 11/9/2025

# Masterflex Pump Annual Calibration Maintenance Guide --- ## Introduction Masterflex pumps are precision instruments widely used in laboratories,...

Masterflex Pump Annual Calibration Maintenance Guide


Introduction

Masterflex pumps are precision instruments widely used in laboratories, pharmaceutical manufacturing, food processing, and other industries requiring accurate fluid handling. To maintain optimal performance and ensure compliance with regulatory standards, annual calibration of Masterflex pumps is essential. Calibration verifies that the pump’s flow rate and speed control are within specified tolerances, ensuring process accuracy, product quality, and safety.

Digital Masterflex pump drives offer ±0.25% speed control accuracy; however, flow rate calibration must be performed whenever tubing is changed or annually, whichever occurs first. This guide outlines a comprehensive procedure for annual calibration using volumetric (graduated cylinder) and gravimetric (weight-based) methods, emphasizing traceability, safety, and documentation for regulated environments.


Safety Precautions

Before performing calibration, observe the following safety precautions:

  • Electrical Safety: Ensure the pump is properly grounded and powered off when connecting or disconnecting equipment.
  • Chemical Safety: Use appropriate personal protective equipment (PPE) such as gloves and goggles when handling fluids, especially hazardous or corrosive liquids.
  • Ergonomics: Use proper lifting techniques when handling heavy equipment like scales or fluid containers.
  • Temperature Control: Perform calibration in a controlled environment at 20 ± 2°C to ensure fluid viscosity and density consistency.
  • Spill Management: Have spill containment materials ready to manage accidental leaks or spills.
  • Follow Manufacturer Instructions: Refer to Masterflex pump user manuals for specific warnings and operational guidelines.

Required Equipment List

| Equipment | Specification / Notes | |------------------------|------------------------------------------------| | Graduated Cylinder | Minimum 100 mL capacity, ±0.5 mL accuracy | | Stopwatch or Timer | Resolution ±0.1 seconds | | Analytical Scale | Minimum 0.1 mg readability, calibrated | | Thermometer | Accuracy ±0.5°C, for ambient temperature check | | Calibration Fluid | Water or fluid matching process fluid viscosity | | Data Recording Sheet | For documenting calibration results | | Cleaning Supplies | For tubing and equipment cleaning before test |


Calibration Procedure Table

| Step | Description | Equipment Used | Acceptance Criteria | |-------|-----------------------------------|-------------------------|-------------------------------------------------| | 1 | Verify ambient temperature (20±2°C) | Thermometer | Temperature within specified range | | 2 | Inspect tubing and pump for damage | Visual inspection | No visible defects or leaks | | 3 | Set pump speed to target flow rate | Masterflex digital drive | Speed control within ±0.25% of setpoint | | 4 | Perform volumetric measurement | Graduated cylinder, stopwatch | Measured volume within ±2% of expected volume | | 5 | Perform gravimetric measurement | Analytical scale, stopwatch | Measured weight within ±2% of expected weight | | 6 | Calculate flow rate and compare | Calculator or software | Flow rate matches pump setpoint ±2% | | 7 | Adjust pump settings if necessary | Masterflex digital drive | Calibration parameters updated and verified | | 8 | Document calibration results | Data recording sheet | Complete and accurate records maintained |


Step-by-Step Calibration Process

Step 1: Preparation

  • Ensure the calibration environment is controlled at 20 ± 2°C.
  • Clean and inspect the pump tubing for wear or damage; replace if necessary.
  • Confirm that the pump is properly installed and connected.
  • Select calibration fluid consistent with normal operation (typically water).

Step 2: Ambient Temperature Verification

  • Use a calibrated thermometer to measure room temperature.
  • Record the temperature; if outside 20 ± 2°C, adjust environment or postpone calibration.

Step 3: Set Pump Speed

  • Power on the Masterflex pump digital drive.
  • Set the pump speed to a target flow rate relevant to typical use (e.g., 100 mL/min).
  • Confirm the speed control reading is within ±0.25% of the setpoint.

Step 4: Volumetric Calibration

  • Place a clean graduated cylinder beneath the pump outlet.
  • Start the pump and stopwatch simultaneously.
  • Collect fluid for a fixed time interval (e.g., 1 minute).
  • Stop the pump and stopwatch simultaneously.
  • Record the volume of fluid collected.
  • Calculate flow rate:
    [ \text{Flow rate (mL/min)} = \frac{\text{Volume collected (mL)}}{\text{Time (min)}} ]
  • Verify the flow rate is within ±2% of the pump’s setpoint.

Step 5: Gravimetric Calibration

  • Tare the analytical scale with an empty container.
  • Collect fluid from the pump outlet into the container for the same fixed time interval.
  • Record the mass of fluid collected.
  • Calculate volumetric flow rate using fluid density (assumed 0.998 g/mL at 20°C for water):
    [ \text{Volume (mL)} = \frac{\text{Mass (g)}}{\text{Density (g/mL)}} ]
  • Calculate flow rate as in Step 4.
  • Confirm flow rate is within ±2% of the setpoint.

Step 6: Adjustment and Verification

  • If flow rates deviate beyond ±2%, adjust pump speed settings accordingly.
  • Repeat volumetric and gravimetric measurements to confirm accuracy.
  • Continue adjustment until acceptance criteria are met.

Step 7: Documentation

  • Record all measurements, environmental conditions, equipment used, and calibration results.
  • Include pump model, serial number, tubing type, calibration date, and technician name.
  • Retain records for traceability and regulatory compliance.

Documentation Requirements

  • Calibration certificates must include:
    • Identification of pump and tubing used
    • Calibration date and technician details
    • Environmental conditions (temperature)
    • Equipment used with calibration status
    • Measurement results and acceptance status
    • Any adjustments made
  • Maintain records in a secure, accessible format for at least 3 years or per regulatory requirements.
  • Ensure traceability to national or international standards (e.g., NIST).

Troubleshooting Calibration Failures

| Issue | Possible Cause | Recommended Action | |-------------------------------|-----------------------------------|---------------------------------------------| | Flow rate below expected | Tubing wear or incorrect size | Inspect and replace tubing; verify tubing size | | Flow rate above expected | Incorrect pump speed setting | Recalibrate speed control; verify settings | | Inconsistent flow readings | Air bubbles or leaks | Purge air from tubing; check for leaks | | Temperature outside range | Environmental control failure | Stabilize room temperature before testing | | Scale or cylinder inaccuracies | Equipment calibration expired | Recalibrate or replace measurement devices |


Frequency Recommendations

  • Annual Calibration: Perform full calibration annually as a minimum.
  • Tubing Change: Perform flow rate calibration immediately after any tubing replacement.
  • After Repairs: Calibrate after any pump maintenance or repair affecting flow.
  • Process Critical Applications: Consider semi-annual or quarterly calibration based on risk assessment.

External Citations

  • Masterflex Pump User Manual, Cole-Parmer Instrument Company, latest edition.
  • ISO 8655-6:2002, Piston-operated volumetric apparatus — Part 6: Gravimetric methods for determination of measurement error.
  • NIST Handbook 44, Specifications, Tolerances, and Other Technical Requirements for Weighing and Measuring Devices.
  • FDA 21 CFR Part 11, Electronic Records; Electronic Signatures — for documentation compliance.
  • Good Laboratory Practice (GLP) Regulations, OECD Principles of GLP.

Legal Disclaimer

This maintenance guide is provided for informational purposes only and does not replace the official Masterflex pump user manual or regulatory requirements specific to your industry or region. Calibration procedures must be performed by qualified personnel using properly maintained and calibrated equipment. The manufacturer and author disclaim any liability for damages or losses resulting from improper use or interpretation of this guide. Always consult official documentation and regulatory bodies for compliance and safety standards.


End of Guide

Was this article helpful?

⚠️ Important Disclaimer

This guide summarizes safe repair practices based on industry best practices and manufacturer guidelines. Always consult the official manufacturer manual for model-specific procedures. Pump equipment repair should only be performed by qualified personnel following proper safety procedures and regulatory requirements.